Which of the following components is primarily used to control the flow of current in an analog circuit?
💡Correct Answer: Option A (Resistor)
In analog electronic circuits, a resistor is the fundamental passive two-terminal component specifically designed to introduce electrical resistance, thereby controlling, reducing, and limiting the flow of electric current according to Ohm's law (I=V/R).
Which of the following is NOT a key optical property of tissues?
💡Correct Answer: Option D (Permeability)
The fundamental optical properties of biological tissue are optical absorption (absorption coefficient μa), optical scattering (scattering coefficient μs and anisotropy factor g), and refractive index/reflection. Permeability is a mechanical or transport property governing fluid and solute flow across membranes, not an optical property.
Which of the following statements is true about an operational amplifier (op-amp)?
💡Correct Answer: Option A (It has a high input impedance and low output impedance)
An ideal operational amplifier possesses infinite input impedance and zero output impedance. In practical op-amps (e.g., IC 741), the input impedance is extremely high (megaohms to gigaohms, drawing negligible current) and the output impedance is very low (tens of ohms), ensuring minimal loading effects.
What is the purpose of feedback in an op-amp circuit?
💡Correct Answer: Option C (To stabilize the output)
Because an open-loop op-amp has an extremely large and unstable open-loop voltage gain (AOL>105) prone to saturation, negative feedback is introduced to stabilize the closed-loop gain, control output voltage, widen system bandwidth, and reduce harmonic distortion.
Which of the following factors is primarily responsible for the absorption of light in biological tissues?
💡Correct Answer: Option C (Melanin and haemoglobin concentration)
In the visible and near-infrared spectral range, the primary endogenous chromophores responsible for optical absorption in biological tissues are melanin (in skin and pigmented tissues) and hemoglobin (both oxyhemoglobin and deoxyhemoglobin in blood).
💡Correct Answer: Option B (To convert one form of energy into another)
A transducer is an electronic device that converts energy from one physical domain into another—most commonly converting non-electrical physical quantities (pressure, sound, temperature, radiation) into proportional electrical signals for measurement and analysis.
💡Correct Answer: Option C (As Low As Reasonably Achievable)
ALARA stands for 'As Low As Reasonably Achievable', a core radiation protection philosophy mandating that all radiation exposures be minimized taking into account economic, societal, and technological factors.
The working principle of a strain gauge transducer is based on:
💡Correct Answer: Option B (Changes in resistance due to strain)
A strain gauge transducer operates on the piezoresistive effect: when mechanical stress or tensile/compressive strain deforms the conductive foil filament, its physical length and cross-sectional area change, producing a proportional fractional change in electrical resistance (ΔR/R=GF⋅ϵ).
Which of the following is used to convert sound energy into electrical signals?
💡Correct Answer: Option B (Microphone)
A microphone is an electroacoustic transducer that detects acoustic sound pressure waves and converts sound mechanical vibrations into proportional electrical signals.
Which of the following is the general form of a complex number?
💡Correct Answer: Option A (a+bi,where a, b in R)
The standard algebraic canonical form of a complex number is z=a+bi (or x+iy), where a is the real part, b is the imaginary part, both a,b∈R, and i=−1.
What is the primary function of the nervous system?
💡Correct Answer: Option B (Coordinate voluntary and involuntary actions and transmit signals)
The primary function of the nervous system is to transmit electro-chemical nerve impulses between different parts of the body, processing sensory inputs and coordinating both voluntary motor movements and involuntary autonomic physiological functions.
Which of the following glands is known as the "master gland" of the endocrine system?
💡Correct Answer: Option B (Pituitary gland)
The pituitary gland (hypophysis) is called the master gland because its anterior and posterior lobes synthesize and secrete trophic hormones (TSH, ACTH, FSH, LH, GH, prolactin) that govern and regulate the functions of other endocrine glands throughout the body.
What is the conjugate of the complex number z = 5-3i?
💡Correct Answer: Option A (5+3i)
The complex conjugate of a complex number z=a−bi is formed by inverting the sign of the imaginary component: zˉ=a+bi. For z=5−3i, the conjugate is zˉ=5+3i.
The residue theorem is primarily used to evaluate integrals of:
💡Correct Answer: Option C (Functions with isolated singularities)
Cauchy's Residue Theorem is a powerful tool in complex analysis used to evaluate closed contour integrals of meromorphic functions containing isolated singular points (poles) via ∮Cf(z)dz=2πi∑Res(f,zk).
An isotope that decreases in activity by 5% per day has an approximate half-life of
💡Correct Answer: Option D (13.5 d)
If activity decreases by 5% per day, the daily remaining fraction is A(1)/A0=0.95. The decay constant is λ=−ln(0.95)≈0.051293 day−1. Thus, the half-life is T1/2=λln2=0.0512930.69315≈13.51 days≈13.5 d.
After the replacement of the ionization chamber in a linac, which machine parameters are expected to change the most?
💡Correct Answer: Option C (Output)
The monitor ionization chamber in a medical linear accelerator directly samples and integrates the beam charge to control beam termination at the designated Monitor Units (MU). Replacing the monitor chamber inevitably shifts the absolute charge collection sensitivity, directly changing the calibrated machine output (cGy/MU), which must be recalibrated before clinical use.
Which of the following is a common biological element used in biosensors?
💡Correct Answer: Option A (Enzymes)
A biosensor combines a biological recognition element with a physicochemical transducer. The most common biological recognition elements are enzymes (e.g., glucose oxidase in glucose biosensors), antibodies, nucleic acids, and whole cells.
________ will absorb the highest dose when exposed to the same fluence of 100-kv photons.
💡Correct Answer: Option A (Bone)
At diagnostic energies around 100 kVp (mean energy ~30-40 keV), the photoelectric effect is the dominant interaction mechanism, whose mass attenuation coefficient scales with (Zeff)3. Cortical bone has an effective atomic number of Zeff≈13.8 compared to water (Zeff≈7.4) and muscle (Zeff≈7.4), causing bone to absorb a significantly higher radiation dose per unit fluence.
As the size of the active volume of the ionization chamber increases, the amount of charge generated for a given dose ________
💡Correct Answer: Option A (Increases)
The total ionization charge Q collected in an ionization chamber is given by Q=D⋅m/(W/e)=D⋅ρV/(W/e). Because the mass of cavity air is directly proportional to the active chamber volume V, a larger volume generates proportionally more charge (higher sensitivity) for a given absorbed dose.
Treatment with a half field can be achieved with either an asymmetric collimator jaw or MLC. Which technique is better?
💡Correct Answer: Option A (Asymmetric collimator)
Asymmetric primary collimator jaws are solid blocks of tungsten (typically 7–8 cm thick) providing transmission of less than 0.1%. In contrast, multileaf collimators (MLC) have rounded leaf ends, tongue-and-groove design, and interleaf leakage with typical transmission of 1.5–2.5%. Therefore, asymmetric jaws produce a significantly cleaner, sharper half-beam block with minimal transmission leakage across the abutment junction.
A patient is set to 100 cm SSD for treatment with an electron beam. One cm of bolus is then added to cover the entire field. The SSD that should be used to calculated the MUs to deliver the desired dose is:
💡Correct Answer: Option C (100 cm)
In electron beam radiotherapy, when bolus is placed on the patient's skin surface, clinical protocol requires setting the patient treatment distance so that the top surface of the bolus is positioned at the nominal calibrated source-to-surface distance (SSD = 100 cm), which matches the machine calibration geometry and cone factor measurements.
If your treatment planning system and treatment delivery system are from different vendors, what DICOM-RT files need to be transferred out of the treatment planning system to effectively treat a patient?
💡Correct Answer: Option A (RT-Plan)
The RT-Plan object contains all geometric and mechanical delivery instructions (beam angles, monitor units, collimator and jaw positions, MLC leaf control points, energy, and accessory codes) required by the linear accelerator and Record & Verify system to execute the patient treatment. RT-Dose and RT-Structure Set are for display/verification, not machine motion execution.
According to the AAPM Task Group 142 report, all of the following are daily QA checks except ________
💡Correct Answer: Option B (Light field/radiation field coincidence)
Under AAPM Task Group 142 recommendations for linac quality assurance, Light Field/Radiation Field Coincidence is classified as a Monthly QA procedure (tolerance 2 mm). Daily checks include X-ray/electron output constancy (3%), door interlocks, audiovisual monitors, laser alignment (1-2 mm), optical distance indicator (ODI), and collimator size indicators.
Which material is commonly used for making biocompatible heart valves?
💡Correct Answer: Option D (Pyrolytic carbon)
Pyrolytic carbon (specifically Low-Temperature Isotropic pyrolytic carbon, LTI carbon) is the biomaterial of choice for mechanical heart valve prostheses due to its exceptional thromboresistance (low blood clotting tendency), hemocompatibility, high fatigue strength, and wear resistance.
What is the primary advantage of using ceramics in dental crowns and bridges?
💡Correct Answer: Option A (High strength and resistance to wear)
Dental ceramics (such as zirconia and alumina-based porcelains) provide superior compressive strength, outstanding abrasion and wear resistance, biocompatibility, and chemical durability in the oral cavity, alongside excellent natural tooth aesthetics.
The Laplace Transform of a function f(t) is given by:
💡Correct Answer: Option A (F(s) = int_0^inf f(t) e^-st dt)
The unilateral Laplace transform of a continuous time-domain function f(t) defined for t≥0 is formally given by F(s)=L{f(t)}=∫0∞f(t)e−stdt, where s=σ+jω is the complex frequency parameter.
The ionization chamber in a radiation therapy system is used to measure:
💡Correct Answer: Option B (The radiation dose delivered to the patient)
In radiotherapy, ionization chambers (such as thimble chambers, Farmer chambers, and plane-parallel chambers) are clinical dosimeters used to quantitatively measure ionization charge in order to determine beam output and the radiation absorbed dose delivered to water/phantom and patients.
Q32
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QABrachytherapy Radionuclides & Seeds
Which of the following radionuclides is most commonly used in prostate brachytherapy due to its appropriate half-life and energy characteristics?
💡Correct Answer: Option A (Iodine-125 (I-125))
Iodine-125 (125I) is the most widely utilized permanent interstitial seed implant radionuclide for low-risk prostate cancer brachytherapy due to its favorable physical half-life (T1/2=59.4 days) and low-energy photon emission (average energy ~28 keV) that spares surrounding rectum and bladder tissues.
What is the region of convergence (ROC) for the Laplace Transform of a signal?
💡Correct Answer: Option A (The values of s for which the Laplace Transform converges to a finite value)
The Region of Convergence (ROC) is the domain of complex frequency values s=σ+jω in the s-plane for which the Laplace transform integral ∫−∞∞∣f(t)e−st∣dt<∞ converges to a finite, bounded value.
Q34
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QAHDR vs LDR Brachytherapy
What is the primary benefit of using a high-dose-rate (HDR) source compared to a low-dose-rate (LDR) source in brachytherapy?
💡Correct Answer: Option A (Reduced time of radiation exposure)
High-Dose-Rate (HDR) brachytherapy delivers radiation at dose rates exceeding 12 Gy/hr (typically using a stepping 192Ir source of ~10 Ci), reducing treatment delivery duration from days (in LDR inpatient hospital stays) to mere minutes, vastly minimizing treatment time and patient discomfort.
Which of the following materials is typically used in joint replacement implants due to its excellent wear resistance and low friction?
💡Correct Answer: Option B (Polyethylene)
Ultra-High-Molecular-Weight Polyethylene (UHMWPE, often highly cross-linked XLPE) is the standard biocompatible polymer used for the bearing articulating cup and spacer inserts in total hip and knee arthroplasty due to its exceptional wear resistance, impact toughness, and low coefficient of friction.
Q36
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QABrachytherapy Radionuclides & Decay
The Ir-192 decay rate is approximately ________% per ________.
💡Correct Answer: Option A (1; day)
Iridium-192 (192Ir) has a radioactive half-life of T1/2=73.83 days. Its decay constant is λ=T1/2ln2=73.830.69315≈0.00939 day−1≈0.94%≈1% per day.
Bremsstrahlung interactions result from electron ________
💡Correct Answer: Option B (Inelastic collisions with atomic nuclei)
Bremsstrahlung ('braking radiation') occurs when a high-speed incident electron undergoes an inelastic Coulomb radiative interaction with the strong electric field of an atomic nucleus, decelerating and deflecting while emitting continuous-spectrum X-ray photons.
The main advantage of flattening filter free (FFF) photon beams is ________
💡Correct Answer: Option B (shorter treatment time)
Removing the metal flattening filter from a linear accelerator increases the beam dose rate by a factor of 2 to 4 (e.g., reaching 1400–2400 MU/min). This dramatically shortens beam-on delivery times, which is particularly beneficial for Stereotactic Body Radiotherapy (SBRT/SRS) and deep inspiration breath-hold (DIBH) techniques.
Complete the decay equation: ^90_38 Sr -> ^90_39 Y + ________
💡Correct Answer: Option A (^0_-1 beta + anti-neutrino)
Strontium-90 undergoes negative beta decay (β− decay) into Yttrium-90: 3890Sr⟶3990Y+β−+νˉe, releasing an electron (−10β) and an electron antineutrino to conserve lepton number and charge.
Which of the following is a directly ionizing particle?
💡Correct Answer: Option B (Electron)
Directly ionizing radiation consists of fast charged particles (electrons, positrons, protons, alpha particles, heavy ions) that deposit energy directly via Coulomb interactions along their tracks. Photons and neutrons are uncharged and therefore classified as indirectly ionizing radiation.
Q41
Diagnostic Radiology Physics, Biomedical Instrumentation & Applied MathematicsDigital Logic & Number Systems
What is the binary representation of the decimal number 13?
💡Correct Answer: Option D (1101)
Converting decimal 13 to binary: 13=8+4+0+1=1⋅23+1⋅22+0⋅21+1⋅20=11012.
Which of the following is/are characteristic(s) of optically stimulated luminescent dosimeters (OSLD)?
💡Correct Answer: Option B (Signal fading exists after irradiation)
Optically Stimulated Luminescent Dosimeters (Al2O3:C) use optical stimulation (green/blue light or laser) rather than heat to release trapped electrons, meaning they can be re-read multiple times without total signal erasure. However, like other solid-state dosimeters, OSLDs exhibit post-irradiation signal fading (~1-2% over the initial 24 hours), which must be accounted for.
Neutrons are produced in high-energy x-ray beam by ________
💡Correct Answer: Option C (Photodisintegration)
Photoneutrons are produced in high-energy medical linacs (energies ≥10 MV) via photodisintegration reactions ((γ,n) and (γ,2n)) when high-energy bremsstrahlung photons exceed the nuclear binding energy threshold of linac target, flattening filter, and collimator materials (lead, tungsten).
Which of the following functions does NOT have a Laplace Transform?
💡Correct Answer: Option A (F(t) = -e^(at))
For a function to have a Laplace transform, it must be piecewise continuous and of exponential order (∣f(t)∣≤Mect as t→∞). While standard power, trigonometric, and delta functions have standard transforms, functions growing faster than exponential order (such as et2, which is frequently typeset or transcribed in competitive exams as −eat or similar non-standard forms) lack a Laplace transform.
4D CT simulation is most useful for treatment planning of tumors located near the
💡Correct Answer: Option C (Diaphragm)
Four-Dimensional Computed Tomography (4D-CT) sorts CT projections into specific respiratory breathing phases, which is clinically vital for imaging and tracking tumors subject to significant respiratory motion—specifically lung, liver, and lower esophageal tumors located near the moving diaphragm.
Which of the following logic gates has the truth table where the output is true only when at least one input is true?
💡Correct Answer: Option B (OR gate)
An OR gate performs logical disjunction: its output is TRUE (1) whenever at least one of its inputs is TRUE (i.e. if either input is 1, or both are 1).
💡Correct Answer: Option A (Excite fluorescent dyes to identify abnormal tissues or cells)
In biomedical fluorescence imaging, monochromatic laser light tuned to a specific absorption wavelength is directed at tissue to optically excite endogenous fluorophores or exogenous fluorescent contrast agents (such as indocyanine green or fluorescein), inducing fluorescence emission to delineate tumors and abnormal cells.
Which of the following is a key advantage of using Monte Carlo simulations for electron transport in radiation dosimetry?
💡Correct Answer: Option B (They can model complex geometries and tissue inhomogeneities accurately)
Monte Carlo simulation is considered the gold standard in radiotherapy dosimetry because it tracks millions of individual particle histories using stochastic interaction cross-sections, providing unmatched accuracy in modeling complex 3D patient geometries, air cavities, bone interfaces, and tissue inhomogeneities.
Monte Carlo simulations in electron transport are especially beneficial in treatment planning for:
💡Correct Answer: Option B (Complex geometries with varying tissue densities and inhomogeneities)
Because electron beam scattering and stopping powers depend heavily on electron density and atomic composition, Monte Carlo simulations are essential for clinical treatment planning in regions with sharp density gradients and tissue inhomogeneities, such as the head-and-neck, lungs, ribs, and nasal cavities.
Q50
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QANuclear Medicine Imaging (PET & SPECT)
Which display system is commonly used to visualize data from Positron Emission Tomography (PET) scans in nuclear medicine?
💡Correct Answer: Option B (Color-coded intensity display)
In PET imaging, regional radiotracer concentration (such as 18F-FDG Standardized Uptake Value, SUV) is standardly rendered using a color-coded intensity display (color lookup tables / heatmaps) to visually highlight hypermetabolic neoplastic lesions against background anatomy.
What type of imaging modality in nuclear medicine benefits most from the integration of fusion imaging on display systems?
💡Correct Answer: Option C (PET/CT)
Hybrid PET/CT systems benefit profoundly from image fusion display: the high functional/metabolic sensitivity of PET is coregistered and overlaid on the high-resolution anatomical cross-sections of CT, allowing precise anatomical localization of hypermetabolic tumors.
What is the term used to describe the amount of radiation delivered by a brachytherapy source over time?
💡Correct Answer: Option A (Dose rate)
Dose rate (expressed in Gy/hr, cGy/min, or cGy/hr) quantifies the amount of absorbed radiation dose delivered by a brachytherapy radioactive source per unit of elapsed time.
Q53
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QABrachytherapy Radionuclides & Seeds
For which type of brachytherapy treatment is the use of a "seed implant" most common?
💡Correct Answer: Option A (Prostate cancer)
Permanent interstitial seed implantation (using encapsulated 125I or 103Pd titanium seeds placed transperineally under ultrasound guidance) is the classic and most common brachytherapy treatment for localized low- and intermediate-risk prostate cancer.
Ionizing radiation primarily damages DNA through the formation of:
💡Correct Answer: Option A (Free radicals)
For low-LET radiation (photons and electrons), approximately two-thirds of cellular radiation damage is mediated by indirect action: the radiolysis of cellular water creates reactive oxygen species and free radicals (primarily hydroxyl radicals, ⋅OH) that diffuse and break DNA phosphodiester backbones.
Which type of DNA damage caused by ionizing radiation is more likely to result in cancer?
💡Correct Answer: Option B (Chromosomal fragmentation or translocations)
While single-strand breaks and base damages are readily repaired by cellular enzymatic pathways, complex double-strand breaks (DSBs) can lead to chromosomal fragmentation, deletions, and non-homologous translocations, which are the primary genetic drivers of radiation-induced genomic instability and carcinogenesis.
Which of the following tissues is most sensitive to ionizing radiation?
💡Correct Answer: Option A (Bone marrow)
According to the Law of Bergonié and Tribondeau, the radiosensitivity of a tissue is directly proportional to its mitotic activity and inversely proportional to its degree of differentiation. Bone marrow (hematopoietic stem cells) has one of the highest cell turnover rates in the human body, making it the most radiosensitive tissue among the options.
Q57
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QARadiopharmaceuticals & Nuclear Medicine
What type of radiation does Technetium-99m emit?
💡Correct Answer: Option C (Gamma rays)
Technetium-99m (99mTc) is an isomeric state that decays via internal transition with a 6.01-hour physical half-life, emitting clean, monochromatic 140.5 keV gamma rays with negligible particulate emission, making it the optimal radionuclide for gamma cameras and SPECT imaging.
Q58
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QARadiopharmaceuticals & Nuclear Medicine
Which of the following radiopharmaceuticals is used for thyroid imaging?
💡Correct Answer: Option A (Iodine-131)
Iodine-131 (131I, as well as 123I and 99mTc-pertechnetate) is standardly used for thyroid scintigraphy, uptake measurements, and targeted radiometabolic treatment of hyperthyroidism and differentiated thyroid cancer.
Which of the following tissues will reflect ultrasound waves the most strongly?
💡Correct Answer: Option D (Air-filled spaces)
The intensity reflection coefficient at an interface depends on the mismatch in acoustic impedance (Z): R=((Z2−Z1)/(Z2+Z1))2. Soft tissue has an impedance of ∼1.63×106 Rayls, whereas air has an impedance of only ∼400 Rayls. This massive acoustic impedance mismatch causes more than 99.9% of the ultrasound energy to be reflected at air-tissue interfaces.
In medical ultrasound, the primary role of signal processing is to:
💡Correct Answer: Option B (Enhance the quality of the ultrasound image by filtering and amplifying the received signals)
The signal processing chain in diagnostic ultrasound takes weak RF echo signals from the beamformer and performs amplification, time-gain compensation (TGC), bandpass filtering, logarithmic compression, envelope detection, and noise rejection to enhance image resolution and contrast.
The penumbra for proton beams ________ as the depth increases.
💡Correct Answer: Option A (Increases)
As protons penetrate matter, they undergo continuous small-angle Coulomb scattering events off atomic nuclei (multiple Coulomb scattering). This cumulative lateral deflection increases with depth, causing the lateral beam edge penumbra to progressively widen as the beam approaches the Bragg peak.
Q62
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QAHDR Brachytherapy Source Calibration
An Ir-192 HDR source is changed between a patient's second and third brachytherapy fractions. Which of the following parameters will be different for the third fractions?
💡Correct Answer: Option C (Treatment time)
Because the new 192Ir source has a significantly higher activity than the old decayed source, the dose rate (D˙) inside the patient increases. To deliver the identical prescribed radiation dose (D=D˙⋅t), the dwell times and total treatment time must decrease inversely with the new source activity.
Q63
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QAAAPM TG-43 Brachytherapy Formalism
An advantage of TG-143 dose calculation is ________
💡Correct Answer: Option D (None of the above is true)
In the standard AAPM brachytherapy dosimetry formalism (TG-43 / TG-143): (1) all calculations assume an unbounded homogeneous water medium without tissue heterogeneity corrections; (2) inter-seed attenuation and shielding effects are ignored; and (3) radiobiological dose-rate variations are not modeled. Therefore, none of the listed statements is an advantage or feature of the TG-43/143 protocol. Choice D.
Q64
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QARadionuclide Therapy (Alpha Emitters)
A radionuclide injection used to treat bone metastasis from prostate cancer is
💡Correct Answer: Option C (Ra-223)
Radium-223 dichloride (223Ra, trade name Xofigo) is an alpha-emitting targeted calcium mimetic that selectively targets areas of high bone turnover in osteoblastic bone metastases from castration-resistant prostate cancer, delivering localized high-LET radiation.
________ cGy is the typical MV-CBCT dose to isocenter for a pelvic case.
💡Correct Answer: Option D (10)
Megavoltage Cone-Beam CT (MV-CBCT) uses beam pulses from the treatment linac (typically 2 to 15 monitor units), delivering an imaging dose of approximately 5 to 15 cGy (nominally ~10 cGy) to the pelvic isocenter per scan, which is higher than kilovoltage CBCT (~1 to 3 cGy).
According to the report of AAPM Task Group 142, the test of cone-beam CT constancy of Hounsfield Units should be performed ________
💡Correct Answer: Option C (Monthly)
In the AAPM Task Group 142 report on quality assurance of medical accelerators (Table VI: Respiratory gating and IGRT), the check for CBCT geometric distortion, spatial resolution, and Hounsfield Unit (HU) constancy is specified as a Monthly quality assurance check.
Tungsten eye shields effectively protect the lens and eye for ________
💡Correct Answer: Option C (Megavoltage electron teletherapy)
Tungsten internal eye shields (lined with low-Z aluminum or plastic to absorb backscattered electrons) are specifically designed to protect the ocular lens and cornea during megavoltage electron beam radiotherapy (e.g., 6–9 MeV electrons) of the eyelid and periorbital lesions. For megavoltage X-rays, several inches of tungsten would be needed, which is clinically impossible to place inside the eye.
The wedge transmission factor(WTF) is 0.90. The MU setting for an open beam is 150. The MU setting for the same dose on the central axis when using the wedge would be ________
💡Correct Answer: Option C (167 MU)
Because the physical wedge attenuates a portion of the radiation beam, the monitor units must be increased inversely with the wedge transmission factor: MUwedge=WTFMUopen=0.90150=166.67 MU≈167 MU.
Considering an AP/PA photon dose distribution prescribed to midplane depth, the hot spot ________
💡Correct Answer: Option B (Decreases for increasing beam energy)
In an opposed pair AP/PA beam arrangement prescribed to midplane depth, higher energy photon beams (e.g., 18 MV vs 6 MV) exhibit greater depth penetration and higher percent depth dose at depth. This creates a flatter, more uniform dose profile across the patient thickness, substantially reducing the magnitude of subcutaneous tissue hot spots.
A linac is calibrated to give 1.00 cGy/MU to d_max for 10x10 cm^2 field at 100 cm SSD. How many MU are needed to deliver 200 cGy to depth if the PDD is 50% for a 10x10 cm^2 field at 100 cm SSD?
💡Correct Answer: Option C (400 MU)
Using the standard percent depth dose calculation formula: Dose at depth=MU×Output×(100PDD) 200 cGy=MU×1.00 cGy/MU×0.50 MU=0.50200=400 MU.
The percent depth dose (PDD) for an MV photon beam decreases with increasing ________
💡Correct Answer: Option B (Depth beyond d_max)
Beyond the depth of maximum dose (dmax), absorbed dose attenuates exponentially as photons are absorbed and scattered in the medium. Therefore, PDD decreases monotonically with increasing depth beyond dmax. (PDD increases with larger field sizes, higher energies, and greater SSDs).
The collimator setting is the field length and width as defined at the level of the ________
💡Correct Answer: Option D (Isocenter)
By international radiation therapy convention, the collimator field size setting (W×L) is universally calibrated and defined at the nominal machine isocenter (typically at a source-to-axis distance of 100 cm).
The field size is set to 16.0x16.0 cm^2 on a linac with 100-cm SAD. What is the field size measured on the patient surface if the patient is set to 130-cm SSD?
💡Correct Answer: Option C (20.8 x 20.8 cm^2)
Radiation beams diverge geometrically from a point source. Using the principle of similar triangles: Field Size at 130 cm=Field Size at 100 cm×(SADSSD)=16.0×100130=16.0×1.3=20.8 cm. Thus, the field size at the surface is 20.8×20.8 cm2.
A patient is treated with an isocentric AP/PA setup on a linac. The anterior SSD is 86 cm, and the posterior SSD is 92 cm. The patient's AP/PA separation is ________
💡Correct Answer: Option C (22 cm)
In an isocentric setup, the linac isocenter is at a distance of 100 cm from the radiation source (SAD = 100 cm): - Anterior depth to isocenter = 100−SSDant=100−86=14 cm. - Posterior depth to isocenter = 100−SSDpost=100−92=8 cm. Total AP/PA separation =14 cm+8 cm=22 cm.
Of the following parameters, ________ does NOT influence the physical penumbra.
💡Correct Answer: Option C (Gantry angle)
The physical penumbra (the lateral distance between the 80% and 20% isodose levels) is composed of geometric penumbra (P=s(SSD+d−SDD)/SDD), transmission through collimator edges, and lateral electron transport/scatter (which depends on beam energy and depth). Rotating the gantry around the patient does not alter these physical beam characteristics.
Q76
Radiation Oncology Physics, Linac Dosimetry & Treatment PlanningBremsstrahlung Spectra & Average Energy
The average energy of a linac's 15-MV photon beam at the surface of the patient is
💡Correct Answer: Option B (5 MeV)
In clinical bremsstrahlung X-ray spectra produced by medical linear accelerators, the mean (average) photon energy is approximately one-third of the nominal accelerating peak potential: Eavg≈31Emax=315 MV=5 MeV.
Which of the following is considered a late somatic effect of radiation exposure?
💡Correct Answer: Option C (Cataracts or leukemia)
Late somatic effects of ionizing radiation manifest months to decades after exposure. Cataractogenesis (deterministic late effect with a threshold) and radiation-induced leukemia/solid tumors (stochastic late effects) are classic late effects, whereas erythema, acute burns, and nausea are early/acute radiation effects.
Q78
Radiation Protection, Shielding, Radiobiology & Waste ManagementRadiobiology & Cell Death
What happens to the cell if the DNA damage caused by radiation is not repaired?
💡Correct Answer: Option A (The cell may undergo apoptosis (programmed cell death))
When severe, complex double-strand DNA damage cannot be successfully resolved by cellular repair mechanisms (e.g., non-homologous end joining or homologous recombination), cell-cycle checkpoints activate apoptotic pathways (programmed cell death) or the cell dies attempting division (mitotic catastrophe).
The inherent filtration in a diagnostic x-ray tube comes from ________ and is usually specified in term of ________
💡Correct Answer: Option D (Components in the x-ray tube itself; mm Al equivalent)
Inherent filtration is the permanent attenuation provided by the physical structural components of the X-ray tube assembly itself—including the glass envelope, insulating dielectric oil, and the tube exit window—and is standardized and specified in millimeters of aluminum equivalent (mm Al equivalent, typically 0.5 to 1.0 mm Al).
What is the effect of increasing the kilovolt (kV) in an industrial X-ray exposure?
💡Correct Answer: Option A (Increases the intensity of the X-ray beam, reducing the exposure time)
Increasing the tube voltage (kVp) increases both the penetrability and the total photon intensity of the X-ray beam (intensity scales approximately with kV2). Because more photons penetrate the subject to reach the detector, the required radiographic exposure time is significantly reduced.
Multi-leaf collimators are usually constructed from ________
💡Correct Answer: Option A (tungsten)
Multi-Leaf Collimator (MLC) leaves are fabricated from sintered high-density tungsten alloy (>90% W, density ρ≈17.5–18.5 g/cm3, Z=74) because of tungsten's exceptional radiation stopping power, structural rigidity, and resistance to deformation during rapid motor movement.
What is the primary purpose of using Monte Carlo techniques in photon and neutron transport?
💡Correct Answer: Option C (To model complex physical interactions and transport phenomena in random but statistically predictable way)
Monte Carlo algorithms use pseudo-random sampling of fundamental probability density functions (cross-sections) to simulate individual particle collisions, scattering angles, and energy depositions, enabling the rigorous statistical modeling of complex particle transport phenomena across heterogeneous media.
Pair production is a process by which a photon interacts with ________
💡Correct Answer: Option C (The nuclear Coulomb field)
Pair production occurs when an incident photon with energy greater than the rest mass energy threshold of two electrons (hν≥2mec2=1.022 MeV) interacts with the intense electrostatic Coulomb field of an atomic nucleus, converting completely into an electron-positron pair.
The most suitable tool for measuring the output factor for a low-energy electron 10x10 cm^2 cutout is ________
💡Correct Answer: Option C (A parallel plate ion chamber)
For low-energy clinical electron beams (<10 MeV), major international dosimetry protocols (AAPM TG-51 and IAEA TRS-398) mandate the use of a plane-parallel (parallel-plate) ionization chamber (e.g., Markus or Roos chambers) because its thin entrance window and negligible cavity perturbation minimize depth and electron fluence errors.
Placing a single in-vivo dosimeter at the prescription point for a TBI treatment can verify that ________
💡Correct Answer: Option A (The patient is at, or close to, the intended SSD)
In Total Body Irradiation (TBI) performed at extended distances (e.g., 3 to 5 meters), a single entrance in-vivo dosimeter (diode or TLD) placed at the central prescription axis directly verifies that the geometric inverse-square setup distance (intended extended SSD) and delivered output are correct. Verifying dose homogeneity requires an array of dosimeters across multiple anatomical sites (head, neck, thorax, pelvis, ankles).
Tests to check the coincidence of the kV and MV isocenters could be performed with ________
💡Correct Answer: Option C (A cube with a small bb placed in its center)
Spatial coincidence between the kilovoltage (kV) imaging isocenter and the megavoltage (MV) radiation delivery isocenter is routinely evaluated using a rigid phantom cube containing a radio-opaque central metallic fiducial sphere (ball bearing / BB) in an automated Winston-Lutz style procedure.
Q88
Radiation Oncology Physics, Linac Dosimetry & Treatment PlanningSRS/SBRT Treatment Plan Quality Indices
The conformity index is a measure of normal tissue irradiated to ________ dose, and the gradient index is a measure of ________
💡Correct Answer: Option B (High; dose falloff)
In stereotactic radiotherapy (SRS/SBRT), the Conformity Index (CI) quantifies the tightness with which the prescription high-dose volume conforms to the target volume (reflecting normal tissue irradiated to high dose), while the Gradient Index (GI, e.g., V50%/V100%) measures the steepness of dose falloff outside the target volume.
Q89
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QABrachytherapy Source Calibration & Well Chambers
Routine calibration of brachytherapy sources is usually carried out with a
💡Correct Answer: Option B (Well-type ionization chamber)
Routine clinical calibration of brachytherapy radioactive sources (such as HDR 192Ir sources and LDR seeds) to determine their Reference Air Kerma Rate (SK) is standardly performed using an accredited, traceable re-entrant well-type ionization chamber.
Q90
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QABrachytherapy Decay & Treatment Time
A vaginal cylinder treatment using ^192 Ir is delivered in 400 seconds on Monday. If the patient is treated again on Thursday, the time to deliver the same dose will be ________
💡Correct Answer: Option C (411 s)
The elapsed time between Monday and Thursday is Δt=3 days. With 192Ir half-life T1/2=73.83 days, the source activity decreases by factor e−λΔt=e−73.830.69315×3=e−0.028166≈0.97223. Since treatment time is inversely proportional to activity, the new treatment time is t=0.97223400 s=411.4 s≈411 s.
💡Correct Answer: Option C (Utilizes previously treated plans to generate a model which can predict treatment planning results and guide optimization)
Knowledge-Based Planning (KBP, such as Varian RapidPlan) uses statistical models trained on libraries of previously treated, clinically validated high-quality treatment plans to predict achievable Dose-Volume Histograms (DVHs) and generate individualized optimization objectives for new patients.
The planning target volume (PTV) margin can be reduced by all of the following techniques except ________
💡Correct Answer: Option D (IMRT)
PTV margins account for geometric uncertainties, inter-fraction setup variations, and internal organ motion. Image-guided radiation therapy (IGRT) modalities like 2D kV orthogonal imaging, kV-CBCT, and MV-CBCT directly measure and correct patient positioning, allowing PTV margins to be safely reduced. IMRT is an intensity-modulation delivery technique that conforms dose to a shape, but does not itself reduce geometric setup or motion uncertainties.
In a dose-volume-histogram, the PTV D_100% value represents the ________ dose received by 100%
💡Correct Answer: Option A (Minimum)
In a cumulative Dose-Volume Histogram (DVH), D100% represents the dose value that is received by 100% of the structure volume, which corresponds to the minimum dose absorbed within that volume.
A limitation of linac-mounted orthogonal kV imaging is ________
💡Correct Answer: Option B (A reduced ability to take orthogonal images when the couch is rotated)
A major physical limitation of on-board linac-mounted kV imaging arms is clearance restriction: when the treatment couch is rotated at non-zero yaw angles, the risk of collision between the imaging source/detector panels and the patient/couch limits the ability to obtain standard orthogonal images without realigning the couch.
100 mCi of ^125 I (half-life 60 days) will decay to 70 mCi in ________ days.
💡Correct Answer: Option C (30)
Using the exponential decay law: A(t)=A0e−λt⟹70=100e−λt⟹ln(0.70)=−λt. With λ=60ln2=600.69315≈0.01155 day−1, the elapsed time is t=0.01155−ln(0.70)=0.011550.35667≈30.87 days≈30 days.
An ^192 Ir source from a high dose rate(HDR) afterloader has an exposure rate of 1.75 R/hr at 1m. It is placed at the center of a container of radius 30cm. ________ TVLs of shielding are required to reduce the exposure rate at surface of the container to less than 2 mR/hr
💡Correct Answer: Option C (4)
Step 1: Calculate unshielded exposure rate at container surface (r=30 cm=0.3 m) using the inverse square law: X˙30cm=X˙1m×(0.31.0)2=1.75×11.11=19.44 R/hr=19,444 mR/hr. Step 2: Calculate required attenuation factor to reduce to <2 mR/hr: Attenuation=219,444=9,722. Step 3: Determine number of Tenth-Value Layers (n): 10n≥9,722⟹n=log10(9,722)≈3.988≈4 TVLs.
Which process is typically modelled in Monte Carlo simulations for photon and neutron transport?
💡Correct Answer: Option B (Interaction with various materials, including scattering, absorption, and transmission)
Monte Carlo radiation transport packages (such as EGSnrc, MCNP, and Geant4) comprehensively simulate the stochastic physical interactions of particles with matter, including Compton scattering, photoelectric absorption, pair production, coherent scattering, neutron elastic/inelastic scattering, and transmission.
The maximum photon energy in an X-ray spectrum is determined by:
💡Correct Answer: Option C (The kVp)
According to the Duane-Hunt law, the maximum kinetic energy of the incident bombarding electrons determines the maximum photon energy in a bremsstrahlung continuous X-ray spectrum: Emax=hνmax=e⋅Vpeak=kVp. Filtration and target material affect the spectrum shape and characteristic peaks, but not the cut-off maximum energy.
💡Correct Answer: Option B (0°-90° with respect to the direction of the incident photon)
By conservation of relativistic energy and momentum in Compton scattering, the recoil electron is always propelled forward in the forward hemisphere at an angle ϕ between 0∘ and 90∘ (0≤ϕ≤π/2) relative to the incident photon trajectory. In contrast, the scattered photon can be deflected at any angle from 0∘ to 180∘.
Which of the following is a common technique used in nanofabrication?
💡Correct Answer: Option A (Lithography)
Lithography (including optical photolithography, extreme ultraviolet lithography, and electron-beam lithography) is the cornerstone top-down patterning technology utilized in nanofabrication to print nanoscale features on semiconductor substrates.
Which technique is used to deposit thin films of material on a substrate in nanofabrication?
💡Correct Answer: Option A (Atomic layer deposition(ALD))
Atomic Layer Deposition (ALD) is a vapor-phase chemical thin-film deposition technique based on sequential, self-terminating surface gas reactions, enabling the deposition of conformal thin films with sub-nanometer atomic thickness control.
💡Correct Answer: Option C (Reduce scatter to the film)
Antiscatter grids (composed of alternating radiopaque lead strips and radiolucent interspace material) are placed between the patient and the image receptor to intercept and absorb scattered Compton photons, thereby preventing image contrast degradation.
Under national and international regulatory standards (AERB in India, US Nuclear Regulatory Commission 10 CFR 35), portable radiation survey meters used for radiation safety monitoring must be calibrated at least once every 12 months (annually) and following any instrument repair.
The most chemically active metals have ________ electron(s) on the outer shell
💡Correct Answer: Option B (1)
The most chemically reactive metals are the alkali metals (Group 1 of the periodic table, e.g., Cs, Rb, K, Na), which possess exactly 1 valence electron in their outermost electron shell, easily lost to achieve a stable noble-gas electron configuration.
Q106
Radiation Protection, Shielding, Radiobiology & Waste ManagementRadioactivity Definitions & Units
The activity of a radioactive isotope is defined as:
💡Correct Answer: Option A (-Delta N / Delta t)
Radioactive activity A is defined as the time rate of nuclear disintegrations of a radioactive sample: A=−dtdN≈−ΔtΔN=λN, where the negative sign denotes the decrease in the number of parent radioactive nuclei over time.
Q107
Radiation Protection, Shielding, Radiobiology & Waste ManagementRadioactivity Definitions & Units
Which of the following is not true? 1Bq:
💡Correct Answer: Option C (Is equal to 3.7x10^10 Ci)
The becquerel (Bq) is the SI unit of radioactivity defined as 1 disintegration per second (1 dps). By historical definition, 1 Curie (Ci)=3.7×1010 Bq. Therefore, 1 Bq=3.7×10101 Ci≈2.7×10−11 Ci, making statement C entirely false.
A 3 MeV photon interacts by the pair production process. What is the combined initial kinetic energy of the positron and electron pair?
💡Correct Answer: Option B (1.98 MeV)
In pair production, the threshold energy required to create the rest mass of an electron-positron pair is 2mec2=2×0.511 MeV=1.022 MeV. The excess incident photon energy is converted into the combined kinetic energy of the pair: EK(e−+e+)=hν−2mec2=3.00 MeV−1.022 MeV=1.978 MeV≈1.98 MeV.
The 2.2 MeV betas from ^90 Sr will travel about ________ in tissue and ________ in air.
💡Correct Answer: Option A (1 cm ; 8m)
The maximum range of high-energy beta particles in unit-density tissue is given approximately by the empirical rule Rtissue≈2Emax cm=22.2=1.1 cm≈1 cm. Since the density of air is approximately 1/800 that of tissue, the range in air is roughly Rair≈1 cm×800=800 cm=8 meters.
All of the following are used in x-ray machine room shielding calculations in the United States, except:
💡Correct Answer: Option D (Instantaneous dose rate)
NCRP Report No. 147 ('Structural Shielding Design for Medical X-Ray Imaging Facilities') bases shielding barrier calculations on weekly workload W (mA⋅min/week), use factor U, occupancy factor T, and shielding design goals P (weekly dose limits). Instantaneous dose rate fluctuates dynamically with pulse width and mA and is not used to determine barrier thickness.
Q111
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QAAAPM TG-43 Brachytherapy Formalism
In the calculation of the dose distribution in brachytherapy, what does the "radial dose function" describe?
💡Correct Answer: Option C (The depth dose profile within the tissue)
In the AAPM TG-43 brachytherapy dosimetry formalism, the radial dose function gL(r) accounts for dose falloff along the transverse bisector of the source due to photon attenuation and scatter in the medium, effectively describing the radial depth dose profile in tissue with the inverse-square geometric effect factored out.
If the SSD of a photon beam is increased from 100cm to 150cm, the PDD at 10 cm depth will:
💡Correct Answer: Option B (Increase more than 2%)
Using the Mayneord F-factor: F=(SSD1+dmaxSSD2+dmax)2×(SSD2+dSSD1+d)2. For a 6 MV beam (dmax=1.5 cm,d=10 cm): F=(101.5151.5)2×(160110)2=(1.4926)2×(0.6875)2=2.228×0.4726=1.053. This corresponds to an increase of approximately 5.3%, which is substantially greater than 2%.
The maximum collimator setting on a linac is 40x40 cm at 100cm SAD. To cover a TBI patient 5 ft. 4 in. (160 cm) tall, with no collimator rotation, the patient's midline must be at a minimum distance of ________ cm
💡Correct Answer: Option C (400)
By linear beam divergence: dField Size at d=100 cmCollimator Size at 100 cm⟹d160 cm=100 cm40 cm⟹d=100×40160=400 cm.
Q114
Radiation Oncology Physics, Linac Dosimetry & Treatment PlanningSuperficial X-Rays vs Electron Beams
Regarding treatment with superficial x-rays, compared with low-energy electrons, all of the following are true except:
💡Correct Answer: Option D (Dose to bone is lower)
Statement D is entirely FALSE: in the superficial X-ray energy range (50–150 kVp), photoelectric absorption dominates and scales with Z3. Because bone has a high effective atomic number (Zeff≈13.8), bone absorbs a dose 2 to 4 times HIGHER than adjacent soft tissue. In contrast, megavoltage electron beams exhibit Compton scattering where absorbed dose depends on electron density, resulting in much lower relative bone dose.
Q115
Brachytherapy, Nuclear Medicine Physics & Radiotherapy QABrachytherapy Radionuclides & Decay
If the half-life of ^103 Pd is 17 days, a consignment of seeds will decay to ________ % of their initial activity after one week.
💡Correct Answer: Option C (75)
After t=7 days (one week), with T1/2=17 days, the remaining activity fraction is: A0A(t)=e−17ln2×7=e−170.69315×7=e−0.2854=0.7517≈75%.
In a mammography screen-film cassette, what is the proper order for the x-rays to penetrate and pass into the cassette?
💡Correct Answer: Option A (Film emulsion > Film base > screen phosphor > screen base)
In screen-film mammography, a single-emulsion film is combined with a single back intensifying screen. The incoming X-ray beam first traverses the front cassette cover, passes through the film emulsion layer, through the transparent film base, and finally strikes the phosphor screen placed behind the film. This arrangement eliminates parallax blur and light diffusion, maximizing spatial resolution for fine microcalcifications.
What does a positive predictive value(PPV1) of 2% for abnormal screening mammography imply?
💡Correct Answer: Option A (Radiologist calling too many abnormals)
In screening mammography benchmarks (BI-RADS / ACR), PPV1 represents the percentage of screening recalls that ultimately result in a tissue diagnosis of cancer (benchmark range: 4% to 10%). A markedly low PPV1 of only 2% signifies an excessively high recall rate with excessive false-positive findings, meaning the radiologist is interpreting too many benign screening examinations as abnormal.
Q118
Radiation Protection, Shielding, Radiobiology & Waste ManagementTransport Regulations & Transport Index
In the context of transporting radionuclides, what does the term "transport index" (TI) refer to?
💡Correct Answer: Option B (A measure of the radiation level at a specific distance from the package)
Under IAEA and DOT/NRC transport regulations, the Transport Index (TI) is a single dimensionless number assigned to a package containing radioactive material, defined as the maximum radiation dose rate measured in millirem per hour (mrem/hr) at a distance of 1 meter from the external surface of the package.
Which type of radioactive waste requires the most stringent containment and isolation measures?
💡Correct Answer: Option C (High-level waste)
High-Level Waste (HLW)—primarily consisting of irradiated spent nuclear fuel and liquid waste generated by reprocessing—contains high concentrations of both short- and long-lived radionuclides and generates substantial decay heat, necessitating deep geological disposal and the most stringent multi-barrier containment for thousands of years.
The biological effects of internal radiation exposure depend primarily on:
💡Correct Answer: Option B (The amount of radiation absorbed by the tissues)
In internal radiation dosimetry, although radiation weighting factors, biokinetics, and route of intake influence deposition, the primary physical determinant of cellular biological effect and tissue morbidity is the total absorbed radiation dose (energy absorbed per unit mass of tissue, measured in Gray / Rad) delivered to the target organs.
🎯 Ready for a Timed Mock Test?
Simulate the real exam experience with a countdown timer, negative marking, and instant detailed scorecards.
About JKSSB Medical Physics Technician 2025 Previous Year Paper
This page provides the full solved question paper for the JKSSB Medical Physics Technician examination conducted in 2025. Every MCQ is presented with verified answer keys and detailed bilingual explanations to support concept building.
Key subjects covered in this paper include Diagnostic Radiology Physics, Biomedical Instrumentation & Applied Mathematics, Radiation Protection, Shielding, Radiobiology & Waste Management, Radiation Oncology Physics, Linac Dosimetry & Treatment Planning, Brachytherapy, Nuclear Medicine Physics & Radiotherapy QA. Practicing authentic previous year questions is the proven way to master question patterns and improve accuracy for upcoming exams across Jammu & Kashmir.